Kowa Saw SL 15 User Manual

PORTABLE SLIT-LAMP  
INSTRUCTION MANUAL  
Kowa SL-15  
 
WARNING  
If any abnormal smell or sound, or overheating or smoke should be detected, turn off the instrument immediately  
and then unplug it.  
If it should continue in use, a fire may break out on the instrument resulting in its malfunctioning. Contact the  
Kowa dealer where you purchased it or your nearest repair shop for inspection.  
Unplug  
When replacing the fuse, make sure the instrument is turned off and unplugged from the socket. If the fuse  
holder cover is removed with the instrument unplugged, there may occur electrical shock.  
Unplug  
Make sure that the power plug or AC adapter is properly inserted into the outlet.  
If not, there may occur a fire or electrical shock.  
Obligatory  
Obligatory  
Use a fuse provided as accessories or a designated one.  
Otherwise, the instrument may malfunction or a fire may break out.  
Make sure that the instrument is properly grounded to protect the body.  
Put the plug in the three-core socket with ground wire.  
Otherwise, there may occur electrical shock.  
Obligatory  
Prohibitory  
Install at a location away from, for instance, a cup containing liquid. If liquid should be spilled into the instrument,  
there may occur electrical shock. If so, turn off the instrument and then unplug it from the outlet. Contact the  
Kowa dealer where you purchased it or your nearest repair shop for inspection.  
Do not disassemble, modify or repair the instrument yourself.  
Otherwise, there may occur a fire, electrical shock, instrument malfunctioning or any injury to the body.  
Contact your Kowa dealer where you purchased the instrument for repair.  
The product assembled by yourself will not get warranty or any other service.  
Disassembly prohibited  
Prohibitory  
The outlet or plug board must not be loaded in excess of its rated capacity.  
If the main power cord should share an outlet with many other devices, there may occur a fire or electrical shock.  
Inserting or guiding into the interior of any metallic object through an air vent slot or opening may cause a fire or  
electrical shock resulting in malfunctioning.  
Prohibitory  
Obligatory  
Do not damage the power supply cable.  
If the power supply cable is damaged,it becomes the cause of a fire or an electric shock.  
Do not modify the power supply cable.  
Do not carry a heavy thing on the power supply cable and do not pull it.  
Do not bring the power supply cable close to a heat equipment and do not make it to overheat.  
Use a lamp provided as accessories or a designated one.  
Otherwise, there may result malfunctioning or a fire.  
Obligatory  
Prohibitory  
Do not short-circuit the terminals of the main unit's charger or a power pack charger. Keep each charging board  
away from any metallic object such as any clip and nail to avoid a fire or electrical shock.  
P - 2  
 
Caution  
Caution  
For unplugging from the outlet, pull the power cord out of it while holding the plug with your hands.  
Otherwise, there may occur a fire or electrical shock.  
Prohibitory  
Prohibitory  
Do not put in, or pull out of, the outlet, the plug with your wet hands to avoid electrical shock.  
Do not install the instrument unstable locations, for instance, on a shaky base or a tilting surface. Otherwise, if it  
should drop off or fall over, the body may be injured.  
Prohibitory  
Prohibitory  
Prohibitory  
Do not replace the illumination lamp by the new one immediately after its service.  
Otherwise, you may burn your fingers due to its high temperature. Allow 30 minutes or so to cool it off before  
replacement.  
The air vent must not be obstructed. Such obstruction will cause internal temperature to increase resulting in  
occurrences of breakdown, malfunction or a fire.  
When in use, keep the instrument away from any high-level source of moisture, dust, soot or steam.  
Otherwise, there may result a fire or electrical shock.  
Prohibitory  
Prohibitory  
Do not touch the instrument or power pack with your wet hands.  
Otherwise, there may occur electrical shock.  
If there is no need to use the instrument for a long period of time, unplug the power cable and power pack for  
storage. Otherwise, there may result a fire.  
Unplug  
Properly mount the power pack in place.  
Otherwise, if it should accidentally drop off, you may get injured.  
Prohibitory  
Do not touch by hand a power pack while charging it, or the power source for a long time, because its temperature  
may increase. If the skin should keep touching the power pack or power source for a long time, it may get burned  
at a relatively low temperature though.  
Prohibitory  
Prohibitory  
Keep your fingers off the gap between the rotary arm and the coupled arm.  
Otherwise, the fingers may be pinched and injured.  
Keep your fingers off the gap around the prism box.  
Otherwise, when adjusting eye-to-eye distance, they may be pinched and injured.  
Prohibitory  
Prohibitory  
Before operating or carrying it around with you, fix the main unit holding it with your hand securely.  
Otherwise, if it should accidentally drop off, you may get injured.  
Do not keep the lamp operating for a long time to avoid a burn your may suffer from a hot lamp-house.  
Prohibitory  
Prohibitory  
When operating the forehead rest fixed with locking screws, tightly fasten them beforehand, to keep them from  
slackening during operation.  
Otherwise, the forehead rest may accidentally jerk toward you to get you injured.  
P - 3  
 
Operational precautions for safety and accident prevention in use  
of battery  
Kowa's SL-15 uses a lithium ion battery in its power pack. Always take the following precautions to avoid a serious injury  
or burn caused by fluid leakage, heat build-up, ignition and explosion as well as occurrence of a fire.  
If any irregularity or fluid leakage should occur:  
G Immediately move the battery further away from fire. Otherwise, leaked fluid or gas may catch fire and explode.  
G If battery fluid got into your eye, rinse it off with clean water and see your doctor at once.  
G If your body or clothes should be contaminated with battery fluid, wash it out.  
Do not put into fire, short-circuit or disassemble the battery. Do not heat it in a microwave oven or conventional  
oven. Do not carry with you, or store, the battery together with metallic objects such as coin, hairpin and necklace  
to avoid short-circuiting.  
Prohibitory  
Do not leave behind, or charge, the battery, for instance, near a fire, under a burning sun or in a car reaching a high  
temperature.  
Prohibitory  
Do not charge the power pack with an other than designated battery charger.  
Do not charge an other than designated power pack.  
Prohibitory  
Do not use or charge, the power battery wet with fluid such as freshwater, salt water, milk, refreshing drinks and  
soapsuds.  
Prohibitory  
Because the protective circuit for preventing any danger inside of the power pack may accidentally break down,  
do not leave behind, or charge, the power pack, for instance, near a fire, under a burning sun or in a car reaching  
a high temperature.  
Prohibitory  
Do not shock the power pack hard, for instance, knocking with a hammer, stamping underfoot or letting it drop  
off.  
Prohibitory  
Prohibitory  
Do not moisten the power pack with fluid such as fresh water, salt water, refreshing drinks and.  
Mount the power pack on the instrument or battery charger, in a proper direction.  
Obligatory  
Obligatory  
If , when it is exhausted, there is no need to use it for a long period of time, remove it from the instrument.  
Recommended practice of battery disposal:  
A lithium ion battery can be recycled. If a power battery you have will not be used any longer, it is  
recommendable to cover its contacts with insulating tape (for instance, cellophane tape) before tak-  
ing it to Kowa, Kowa's dealer where you purchased it or any recycling operator.  
P - 4  
 
Operational precautions  
G The instrument should be installed and/or stored in a dust free place free from high temperatures, high humidity and  
direct sunlight. The environmental conditions described below should be observed strictly.  
In operation  
10˚C to 40˚C  
30% to 75%  
In storage  
-15˚C to 60˚C  
10% to 95%  
Environmental temperature  
Relative humidity  
G Be careful about where the instrument is to be installed, to keep it from accidentally being unplugged during opera-  
tion. If it should be unplugged, be sure to turn off the power switch before plugging again.  
[ Operation ]  
G Handle the instrument carefully so that the instrument will not be shocked hard. When the instrument was shocked  
hard, dropped off or damaged in any degree, do not operate the power source or the main unit before taking mea-  
sures.  
G When the main unit is not in use, be sure to always cover the instrument for protection.  
G Set all the dials and knobs with click(s) at their proper click positions.  
G Do not operate lamp .continuously. Make it a rule to operate it for about two minutes immediately followed by at least  
three minutes of turn-off.  
[ Power source/charging ]  
G Charge a designated power pack only, to avoid a danger of trouble. Any other battery may explode to injure the body  
and articles around it.  
G The line voltage should be within 10% of the rated value.  
G Do not turn on and off the power switch consecutively, but at an interval of at least five seconds instead.  
G Plug and unplug with the power switch off.  
[ Cautions in miscellaneous matters ]  
G Kowa is not responsible for product failure and damage caused by modification, maintenance or repair rendered by  
an other than designated service operator.  
G Kowa is not responsible for product failure and damage caused by modification, maintenance or repair rendered  
using other than designated parts as expressed and implied in the warranty certificate.  
G Since this instrument uses precision parts which will require special tool for adjustment, never disassemble or adjust  
it yourself.  
P - 5  
 
Operational Considerations for Hospital Grade Electrical  
Equipment (Safety and Accident Prevention)  
This section describes general precautions concerning the use of medical electrical instrument based on Notification No.  
495 of the Pharmaceutical Affairs Bureau of the Ministry of Health and Welfare dated June 1, 1972.  
1. Only qualified personnel should operate this instrument.  
2. The following items shall be considered when installing instrument.  
(1) Install at a location away from water or accidental splashing.  
(2) Install at a location which will not be adversely affected by atmospheric pressure, temperature, humidity, ventilation,  
sunlight, dust, air containing salt, sulfur and other substances, and the like.  
(3) Take care to guard against tilt, vibration and strong impacts, for instance, during transportation.  
(4) Equipment must not be installed at locations where chemicals are stored or gasses are generated.  
(5) Be careful with the radio frequencies, voltages and allowable amperes (power consumption) of the power supply.  
(6) Make sure that all batteries are installed properly and in good working order (discharging curve, polarity, and so forth).  
(7) Properly connect ground wires.  
3. The following items shall be considered when using the instrument.  
(1) Make sure that instrument activates properly after checking switch contact. Polarity, dial setting and meters and so forth  
(2) Make sure that the instrument is properly grounded.  
(3) Make sure that all cords are properly connected and secured.  
(4) Use of other instruments and appliances on the same power circuit is liable to cause errors and incorrect flash output  
resulting in incorrect diagnosis or hazards.  
(5) External circuits and connectors that may come in direct contact with the patient must be checked frequently for signs  
of wear.  
(6) Before operations, make sure that the camera back battery is sufficiently charged.  
4. The following items shall be considered when using the instrument.  
(1) Be sure to minimize the time and quantity required for diagnosis and treatment.  
(2) Always assure that the instrument and patient are in good condition.  
(3) When an abnormality is found on the instrument, take proper measures, for instance, to stop the operation of the  
instrument while assuring the patient's safety.  
(4) Do not allow the patient to touch any of the instrument controls.  
5. The following items shall be considered after using the instrument.  
(1) Turn off the instrument after setting control switches, dials and so forth to their initial statuses following with a specified  
procedure.  
(2) Do not pull cords for removal because an excessive force is exerted on them.  
(3) The following shall be considered regarding storage location.  
(a) Store the instrument at locations free from splashes of water.  
(b) Store at a location which will not be adversely affected by atmospheric pressure, temperature, humidity, ventilation,  
sunlight, dust, air containing salt, sulfur and other substances, and the like.  
(c) Take care to guard against tilt, vibration and strong impacts, for instance, during transportation.  
(d) Equipment must not be stored at locations where chemicals are stored or gasses are generated.  
(4) Clean and rearrange accessories, cords, cord restraints and the like.  
(5) The instrument must be cleaned prior to use so that there will be no problem when using it again.  
6. If it has some trouble, a label describing the trouble should be affixed on the instrument and contact a  
repair shop for repair.  
7. Equipment shall not be modified.  
8. Maintenance  
(1) Periodically check the instrument and its components for any abnormality.  
(2) When using again the instrument that has not been used for a while, it must be checked beforehand to assure that it is  
in normal condition and operates safely.  
P - 6  
 
Accessories  
Power cable:1 pc.  
Fuse:2 pc.  
Lump:1 pc  
Dust cover:1 pc.  
Instruction manual:1 pc.  
P - 7  
 
Contents  
Introduction ............................................................................................................. 1  
Operational Considerations for Safety and Accident Prevention ............................ 1  
Precautions in use of battery (safety/accident prevention  
and recommended practices of battery disposal) .................................................... 4  
Operational precautions........................................................................................... 5  
Operational Considerations for Hospital Grade Electrical Equipment  
(Safety and Accident Prevention) ............................................................................ 6  
Accessories ............................................................................................................ 7  
Contents .................................................................................................................. 8  
1 Name and function of each part ...................... 9  
2 Preparation ..................................................... 12  
2-1 Setting ........................................................................................................ 12  
2-2 Charging the power pack inside of the main unit ........................................ 12  
2-3 Charging a single power pack independently of the main unit .................... 13  
2-4 How to fit an optional items (forehead rest)................................................. 13  
2-5 Charge time and service time of power pack .............................................. 14  
3 How to use the power pack ........................... 15  
3-1 How to hold slit-lamp .................................................................................. 15  
3-2 Adjusting each part...................................................................................... 15  
3-3 How to use magnification selector, slit disk and filter disk .......................... 16  
4 Troubleshooting............................................... 17  
5 Maintenance and inspection........................... 18  
5-1 Daily inspection ........................................................................................... 18  
5-2 Replacing lamp............................................................................................ 18  
5-3 Replacing fuse............................................................................................. 19  
6 Specifications.................................................. 20  
P - 8  
 
Name and function of each part  
1.  
1 Prism box  
Main unit (slit-lamp)  
For adjusting the eye-to-eye distance, bend  
and stretch this part both ways.  
2 Ocular  
With rubber attachment for compen-  
sating the diopter of the laterally bi-  
ased eye either way.  
3 Objective lens  
6 Main lever (ON/OFF select switch)  
An image 10 times as large as a subject  
can be recorded.  
An image 16 times as large as a subject  
can be recorded by advancing the lens.  
Press the lever with its grip being held by hand, to  
energize the lamp.  
Release the lever to turn it off.  
7 Upper cover  
Remove the insert-type cap in the center to fit  
optional items.  
5 Slit-disk  
For selecting a desired slit  
width from the range of spot,  
0.1 mm, 0.2 mm and 0.8 mm.  
D Magnifying power  
select lever  
Turn the lever in the direc-  
tions of  
to select a mag-  
nifying power of 16 or 10.  
Magnifying powers in be-  
tween the range cannot be  
selected.  
8 Filter disk  
For selecting light intensity  
and use/nonuse of blue filter.  
Select a desired light inten-  
sity from the range of full, 1/  
4 and 1/16.  
9 Grip cover  
For covering the built-in  
power pack inside of the in-  
strument  
C Lamp cap  
To be removed when replac-  
ing a lamp, which is located on  
the left-hand side as viewed by  
the examiner.  
0 Coupled arm  
To be frontally inserted into the  
power source's receptacle for  
charging the main unit  
A Grip  
4 Angular graduations of arm's swing  
For indicating the angle of illumination lamp's  
swing  
B Lamphouse  
For housing a lamp inside. Do not touch  
it bare-handed because it reaches a high  
temperature.  
P - 9  
 
1. Name and function of each part  
Picture of power source as viewed from above  
F Main unit's charger indicating lamp  
When the power pack inside of the main unit, which is set  
at the main unit's charger, is being charged, the indicating  
lamp comes on to emit orange light. When charging is  
complete, it comes on to emit green light.  
G Main unit's charger (also serving as a base  
for holding slit-lamp)  
With the power switch turned on, push the slit-lamp deep  
into the receptacle, to enable the power pack inside of  
the main unit, to be charged.  
E Power source indicating lamp  
When the power switch is turned on, the  
indicating lamp comes on to emit blue light.  
H Inlet  
To be connected to the power cable  
I Power switch  
For turning on/off the  
power source.  
(The figure seen from the top)  
J Power pack charger  
For charging the power pack taken out from the main unit  
or a spare power pack.  
K Power pack charger indicating lamp  
When the power pack set at the power pack  
charger is being charged, the indicating lamp  
comes on to emit orange light. When charging is  
complete, the indicating lamp comes on to emit  
green light.  
(The figure seen from the bottom)  
L Fuseholder  
Replace a blown-out fuse by a designated new one.  
P - 10  
 
1. Name and function of each part  
Battery (power pack)  
M Battery connector  
When it is set at the main unit or power pack charger, set the  
battery connector such that it can be in proper contact with the  
charger.  
Option (forehead rest)  
N Forehead rest  
Fix the forehead rest such that the examined  
patient's forehead can rest against it.  
Q Fixed plate  
Fit the fixed plate on the main unit.  
P Arm  
O Locking screw  
The arm will expand or contract de-  
pending on how hard the forehead  
rest is pressed.  
Arm's motion can be fixed to a de-  
sired extent depending on how tight  
the arm is fastened with the locking  
screws.  
P - 11  
 
Preparation  
2.  
2-1 Setting  
Place the power source on a stable table and connect the power cable to the inlet on the back of the power source.  
Connecting the power cable  
2-2 Charging the power pack inside of the main unit  
1 Open the grip cover by pulling it out toward you while depressing the rubber portion of grip cover downward.  
2 Make sure the power pack is orientated in a right direction before mounting it on the main unit.  
3 Hook the claw located at the lower portion of grip cover to close the grip cover.  
1 Open the grip cover  
2 Mount the power pack  
3 Close the grip cover  
4 Insert the power cable deep into the inlet at the back of power source while holding the main unit tightly.  
5 Turn on the power switch to start charging.  
Make sure the power source indicating lamp is on emitting blue light and the main unit's charger indicating lamp is on  
emitting orange light.  
When charging is complete, light which the indicating lamp emits will switch to a green one.  
After inserting the power cable into the main unit, make sure the indicating lamps are on emitting light of specified  
color.  
4 Insert the main unit  
5
Make sure the indicating the lamp works normally.  
P - 12  
 
2-3 Charging a single power pack independently of the main unit  
2-3 Charging a single power pack independently of the main unit  
1 Insert the power pack deep into the power pack charger while holding it in the direction of photo.  
2 Make sure the power pack charger indicating lamp is on emitting orange light. When charging is complete, light which  
the indicating lamp emits will switch to a green one.  
1 Insert the power pack.  
2
Make sure the indicating lamp works normally.  
Do not short-circuit the terminals of the main unit's charger or a power pack charger. Keep  
each charging board away from any metallic object such as any clip and nail to avoid a fire  
or electrical shock.  
2-4 How to fit optional item (forehead rest)  
1 Open the cover located on an upper portion of the main unit by a pointed stick.  
2 Tighten screws by an object such as a piece of coin to fit the forehead rest.  
3Rotating and loosening the lock-  
ing screw springs the forehead  
rest back and forth.  
1 Open the cover  
2 Tighten screws.  
When operating the forehead rest fixed with locking screws, tightly fasten them  
beforehand, to keep them from slackening during operation.  
Otherwise, the forehead rest may accidentally jerk toward you to get you injured.  
P - 13  
 
2. Preparation  
2-5 Charge time and service time of power pack  
1 The reference charge time spent in fully charging an exhausted battery at a room temperature of 25˚C is indicated  
below:  
Charge time  
With main unit charger  
About 90 minutes  
About 120 minutes  
With power pack charger  
Caution  
Keep in mind that charge time will vary depending on what condition the power source is placed under, how high its  
ambient temperature is, and how often it is used for charging.  
Generally, charge time will increase with the frequency of use because of power pack's degradation.  
2 Service time of power pack  
Service time  
About 40 minutes at a time  
Do not keep the lamp operating for a long time to avoid a burn your may suffer from a hot  
lamp-house.  
Caution  
Keep in mind that power pack's service time will vary depending on how high its ambient temperature is, and how  
often it was charged.  
Generally, the power pack 's service time will decrease with its frequency of use because of its deterioration.  
3 Power pack's service lifetime (as a standard to determine when the power pack in service should be replaced by a  
new one).  
The power pack's service time after about 500 cycles of charging and discharging at an ambient temperature of 20  
Åé is approximately 20 minutes (as a standard to determine how long the service time will be).  
If the service time seems to be shorter, replace it by a new one.  
Caution  
If, the power pack should be left unused after charging fully, its performance will gradually decrease even if it will not  
be used at all.  
If you have spare power packs, periodically, utilize all your power packs by turns to keep their performance as high  
as possible.  
P - 14  
 
How to use  
3.  
3-1 How to hold the slit-lamp  
1 Hold the slit-lamp's main body. securely grabbing its grip by your right (or left) hand.  
2 Place the other hand's thumb on the side of upper-end of the slit-lamp's main body.  
3 Stabilize slit width indication by adjusting the distance between the examined eye and the objective lens, depending  
on how wide the opening is between the thumb and the fingers when put on the patient's forehead.  
(If optional forehead rest is used, it is much easier to fix the main body.)  
How to hold the slit-lamp  
3-2 Adjusting each part  
1 Adjusting eye-to-eye distance  
Adjust the eye width to the examined eye by binding and stretching both prism boxes both ways while holding them  
by hand, to have an adequately round field of vision as viewed through the ocular lens.  
2 With the ocular lens turned to its full plus(+) range, look into the viewfinder with each eye while slowly turning it  
clockwise. When the line of sight comes into clear sight, stop turning. It is recommended that a patent wearing eye-  
glasses should use the viewfinder with its rubber eye piece collar folded forward.  
1 Adjusting eye-to-eye distance  
2 Adjusting ocular diopter value  
P - 15  
 
3. How to use  
3-3 How to use magnifying power select lever, slit-disk and filter-disk.  
1 Turn the magnifying power select lever clockwise as indicated by arrow-head to select a magnifying power of 16.  
Turn it counterclockwise to select a magnifying power of 10.  
2 Slit-disk  
Turn the slit-disk as far as it is set at one of the four click stops, which will cover three types of slit widths (0.1 mm, 0.2 mm  
and 0.8 mm). The point marking "O" indicates "spot." Slit-width will be indicated at a point marking a red circle "G".  
3 Filter-disk  
For selecting to a desired light intensity from four types (full, 1/4 and 1/16). If "B" is selected, blue filter will be applied.  
1 Magnifying power selection  
2 Switching by disk  
P - 16  
 
Troubleshooting  
4.  
This section describes how to take action against any irregularity which occurred on the instrument. If any irregularity  
should occur on the instrument, check if the following faults are observed, before contacting Kowa's dealer where you  
purchased the instrument for repair.  
Be sure to turn off the power source before repairing.  
G
G
G
G
G
G
G
G
P - 17  
 
Maintenance and inspection  
5.  
5-1 Daily inspection  
1 If you do not operate the instrument any longer for the day, be sure to turn off the power switch and put a cover on it.  
2 Care must be used to keep the objective lens free from dust and dirt, fingerprints, tear-drops and so forth.  
3 Clean fouled main unit, power source and power pack with a piece of soft cloth. If dirt is stubborn, apply diluted  
detergent to it. Do not apply chemicals or solvents such as thinner and benzene to avoid deterioration, deformation or  
paint peel-off.  
4 If there is no need to use it for a long period of time, remove the power pack from the main unit and unplug the power  
cable from the outlet.  
5-2 Replacing the lamp  
1 Make sure that the lamp has cooled down before replacement.  
2 Remove its cap before removing the lamp.  
3 After fitting a new lamp, replace the cap back in place.  
1 Remove the lamp cap  
2 Replacing the lamp  
Caution  
Insert the lamp deep into its lamp socket for fitting.  
Do not touch the lamp bare-handed to prevent hand's fat from transferring to the lamp surface. Hot lamp surface  
may burn the fat resulting in a shorter lifetime.  
The lamp is exclusive use of SL-15.It can not be used for SL-14.  
Caution  
Caution  
Use a lamp provided as accessories, or a designated one.  
If an other than designated lamp is in use, there may result malfunctioning or occurrence  
of a fire.  
Do not replace the lamp immediately after use.  
Because the lamp in use is scorching hot, you may burn your hand on it.  
Allow thirty minutes or so to cool it off before replacement.  
P - 18  
 
5-3 Replacing fuse  
5-3 Replacing fuse  
1 With the power switch turned off, unplug the power cable from the outlet.  
2 Unscrew the power source's fuse holder cover by turning minus (-) screw driver counterclockwise to take out the fuse  
from within.  
3 Make sure a new fuse to be used, is a designated capacity and type. Then replace the fuse by a new one and tight-  
screw the fuse holder cover.  
Caution  
Caution  
Before replacing the fuse, turn off the power switch and unplug the power cable  
from the outlet.  
Do not remove the fuse holder cover with the power cable unplugged, there may  
occur electrical shock.  
Use a fuse provided as accessories or a designated one.  
If an other than designated fuse should be used, there may result malfunctioning or  
occurrence of a fire.  
P - 19  
 
Specifications  
6.  
Main unit (slit-lamp)  
[Microscope]  
Type :  
Binocular-stereoscopic-orthoscopic microscope  
Inward angle :  
Magnifying power :  
13˚  
10/16  
Objective lens' working distance :  
About 80 mm (when a magnifying power of 16 is selected)  
About 100 mm (when a magnifying power of 10 is selected)  
Ø 10 mm (when a magnifying power of 16 is selected)  
Ø 15 mm (when a magnifying power of 10 is selected)  
Double-eye pieces  
Practical field of view :  
Reticle :  
2-magnifying power selectable/moving objective lens type  
50 mm /70 mm  
Eye-to-eye distance :  
Ocular diopter :  
-8 to + 4D  
[Slit-projector]  
Slit type :  
Turret-type switching  
Slit length :  
12 mm fixed  
Slit width :  
0.1 mm, 0.2 mm, 0.8 mm, and Ø12 mm spot  
Full, 1/4, 1/16 (three stages)  
60° with respect to its horizontal outer periphery  
Built-in blue filter  
Light intensity adjusting :  
Slit's illuminating angle :  
Filter :  
Continuous lamp operating time :  
About 40 minutes  
[Electrical ratings]  
According to the type of protection against electric shock :  
According to the degree of protection against electric shock  
Internally powered equipment  
type B Applied part  
:
Electromagnetic compatibility :  
Line voltage :  
IEC 60601-1-2 (1993)  
8.4 V d.c. (special purpose battery)  
15 W  
Power consumption :  
Main unit-External dimensions/Weight :  
221 mm (W)_106 mm (D)_214 mm (H)/790 grams  
Power source (charger)  
[Type and electrical ratings]  
Type :  
Also servicing as a base for holding a slit-lamp)  
For changing a built-in power pack and a single one.  
100 V a.c. 50 Hz or 60 Hz  
Output terminal :  
Line voltage :  
Line source :  
90 VA  
According to the type of protection against electric shock :  
According to the degree of protection against electric shock :  
Class I equipment  
Nothing  
Electromagnetic compatibility :  
Power source/Weight :  
IEC 60601-1-2 (1993)  
124 mm (W)_253 mm (D)_68 mm (H)/1250 grams  
P - 20  
 
Life Science Division  
No. 4-14, 3-chome, Nihonbashi-Honcho, Chuo-ku, Tokyo 103-8433, Japan  
Phone: 81(3) 3279-7331  
Immermannstrasse 65A  
20001 So. Vermont Ave. Torrance,  
CA 90502, U.S.A.  
40210 Düsseldorf F.R. Germany  
Phone: 49(211) 35-3444/45/46  
Facsimile: 49(211) 161952  
Facsimile: 81(3) 5255-7516  
Phone: 1(310) 327-1913  
Facsimile : 1(310) 327-4177  
Printed on recycled paper.  
Printed in Japan SL 150 V1.1 E 030701 MS  
 

Insignia Automobile Parts NS MB5000 C User Manual
Ives Door SRT User Manual
JBL Portable Speaker DS35 User Manual
Kambrook Hot Beverage Maker KSK65 User Manual
Kenmore Water System 625348460 User Manual
Kodak Digital Camera 1845346 User Manual
Kodak DVD Player DVD 40 User Manual
Korg Recording Equipment EA 1 User Manual
Kramer Electronics Network Hardware RTBUS 21 User Manual
Kramer Electronics Ventilation Hood rk 1utb User Manual